Infection warning over faulty mattress covers

Nurses have been urged to check special thermo contour mattresses that are designed to prevent the occurrence of pressure ulcers because of a potential infection risk, according to an alert from the government’s medicines and healthcare products watchdog.

Due to a material fault, some Thermo contour mattress covers manufactured by Barrington Healthcare International and supplied by Hill-Rom, have aged prematurely and developed small holes, said the Medicines and Healthcare products Regulatory Agency.

This has allowed fluids to pass through and contaminate the mattress’s foam core, posing a potential infection risk to patients.

Staff have been instructed by the MHRA to identify all Thermo contour mattress covers with lot numbers from 3162 to 3626 inclusive.

They should remove the cover and inspect inside the foam core for contamination, said the medical device alert.

Contaminated mattresses and foam cores or damaged covers must be removed from use and replaced and staff should contact Hill-Rom to obtain replacement covers for affected lots

Have your say

You must sign in to make a comment.

Related images

Related Files

Online training units, written and reviewed by experts. Earn two hours' CPD and a personalised certificate for your portfolio.

Subscribers get five FREE learning units and non-subscribers can access each learning unit for £10 + VAT.

Click here to find out more

Related Jobs

Sign in to see the latest jobs relevant to you!